Surgical treatment of tricuspid valve insufficiency promotes early reverse remodeling in patients with axial-flow left ventricular assist devices

J Thorac Cardiovasc Surg. 2012 Jun;143(6):1370-6. doi: 10.1016/j.jtcvs.2011.07.014. Epub 2011 Aug 11.

Abstract

Objective: The HeartMate II (Thoratec Corp, Pleasanton, Calif) continuous-flow left ventricular assist device has emerged as the standard of care for patients with advanced heart failure. The objective of this study was to assess the safety and early effectiveness of concomitant tricuspid valve procedures in patients undergoing implantation of a HeartMate II device.

Methods: From February 2007 to April 2010, 83 patients underwent HeartMate II left ventricular assist device implantation. Of these, 37 patients had concomitant tricuspid valve procedures (32 repairs, 5 replacements) for severe tricuspid regurgitation. The effects of a tricuspid valve procedure on tricuspid regurgitation and right ventricular remodeling were assessed comparing echocardiographic findings at baseline and 30 days after left ventricular assist device implantation. Overall survival was also compared.

Results: Patients undergoing a concomitant tricuspid valve procedure had more tricuspid regurgitation (vena contracta, 5.6 ± 2.1 mm vs 2.9 ± 2.0 mm; P < .001), worse right ventricular dysfunction (right ventricular end-diastolic area, 33.6 ± 6.2 mm vs 31.6 ± 8.5 mm; P = .05), higher mean right atrial pressure (17.4 ± 7.1 mm Hg vs 14.9 ± 5.1 mm Hg; P = .03), and a higher Kormos score (2.6 ± 2.1 vs 1.2 ± 1.4; P = .0008) preoperatively. One month after surgery, tricuspid regurgitation was worse in patients who underwent left ventricular assist device implantation alone (+18.6%), whereas it improved significantly in patients undergoing a concomitant tricuspid valve procedure (-50.2%) (P = .005). A corresponding significant reduction in right ventricular end-diastolic area (33.6% ± 6.2% vs 30.1% ± 9.7%; P = .03) and a trend toward better right ventricular function (55.5% ± 79.7% vs 35.7% ± 60.5%; P = .28) were noted in patients undergoing a concomitant tricuspid valve procedure. Survival was comparable between the 2 groups.

Conclusions: In patients with severe tricuspid regurgitation undergoing left ventricular assist device implantation, a concomitant tricuspid valve procedure effectively reduces tricuspid regurgitation and promotes reverse remodeling of the right ventricle.

MeSH terms

  • Aged
  • Cardiac Valve Annuloplasty* / adverse effects
  • Cardiac Valve Annuloplasty* / mortality
  • Chi-Square Distribution
  • Disease-Free Survival
  • Female
  • Heart Failure / complications
  • Heart Failure / diagnostic imaging
  • Heart Failure / mortality
  • Heart Failure / physiopathology
  • Heart Failure / therapy*
  • Heart Valve Prosthesis Implantation* / adverse effects
  • Heart Valve Prosthesis Implantation* / mortality
  • Heart-Assist Devices*
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Minnesota
  • Prosthesis Design
  • Retrospective Studies
  • Severity of Illness Index
  • Time Factors
  • Treatment Outcome
  • Tricuspid Valve Insufficiency / complications
  • Tricuspid Valve Insufficiency / diagnostic imaging
  • Tricuspid Valve Insufficiency / mortality
  • Tricuspid Valve Insufficiency / physiopathology
  • Tricuspid Valve Insufficiency / surgery*
  • Ultrasonography
  • Ventricular Function, Left*
  • Ventricular Function, Right*
  • Ventricular Remodeling*