Rapid quantification of quinine and its major metabolite (3S)-3-hydroxyquinine in diluted urine by UPLC-MS/MS

J Pharm Biomed Anal. 2011 Jun 1;55(3):494-9. doi: 10.1016/j.jpba.2011.02.004. Epub 2011 Mar 5.

Abstract

A rapid UPLC-MS/MS quantitative assay for the quantification of quinine and (3S)-3-hydroxyquinine requiring minimal sample pre-treatment - dilute-and-shoot type approach - has been developed. The assay was run at 0.6mL/min using gradient elution with (pH 10; 10mM) ammonium bicarbonate and methanol with a total cycle time of 2.5min on a 50mm×2.1mm ID, 1.7μm Acquity BEH column. Peak shapes were highly symmetrical allowing for accurate peak integration. Calibration curves for both analytes were constructed from 1.00 to 20.00ng/mL, yielding R(2) values >0.995. Intra- and inter-batch assay precision and accuracy were evaluated using 6 injections of QC solutions on 3 separate days (n=18) and were found to be within ±10% and 90-110% respectively. The method was shown to be suitable for quantitatively determining the ratio of quinine to (3S)-3-hydroxyquinine for a cohort of samples from an epidemiological study.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Calibration
  • Chromatography, High Pressure Liquid / instrumentation
  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Limit of Detection
  • Quinidine / analogs & derivatives*
  • Quinidine / urine
  • Quinine / urine*
  • Reference Standards
  • Reproducibility of Results
  • Solid Phase Extraction
  • Solutions
  • Tandem Mass Spectrometry