Hemocompatibility assessment of carbonic anhydrase modified hollow fiber membranes for artificial lungs

Artif Organs. 2010 May;34(5):439-42. doi: 10.1111/j.1525-1594.2009.00882.x.

Abstract

Hollow fiber membrane (HFM)-based artificial lungs can require a large blood-contacting membrane surface area to provide adequate gas exchange. However, such a large surface area presents significant challenges to hemocompatibility. One method to improve carbon dioxide (CO(2)) transfer efficiency might be to immobilize carbonic anhydrase (CA) onto the surface of conventional HFMs. By catalyzing the dehydration of bicarbonate in blood, CA has been shown to facilitate diffusion of CO(2) toward the fiber membranes. This study evaluated the impact of surface modifying a commercially available microporous HFM-based artificial lung on fiber blood biocompatibility. A commercial poly(propylene) Celgard HFM surface was coated with a siloxane, grafted with amine groups, and then attached with CA which has been shown to facilitate diffusion of CO(2) toward the fiber membranes. Results following acute ovine blood contact indicated no significant reduction in platelet deposition or activation with the siloxane coating or the siloxane coating with grafted amines relative to base HFMs. However, HFMs with attached CA showed a significant reduction in both platelet deposition and activation compared with all other fiber types. These findings, along with the improved CO(2) transfer observed in CA modified fibers, suggest that its incorporation into HFM design may potentiate the design of a smaller, more biocompatible HFM-based artificial lung.

MeSH terms

  • Animals
  • Carbon Dioxide / metabolism
  • Carbonic Anhydrases / metabolism*
  • Enzymes, Immobilized / metabolism*
  • Heart-Lung Machine*
  • Materials Testing*
  • Membranes, Artificial*
  • Sheep
  • Surface Properties

Substances

  • Enzymes, Immobilized
  • Membranes, Artificial
  • Carbon Dioxide
  • Carbonic Anhydrases