Introduction: We previously demonstrated that wavefront synchronization by spatiotemporal excitable gap pacing (Sync P) is effective at facilitating spontaneous termination of ventricular fibrillation (VF). Therefore, we hypothesized that a spatiotemporally controlled defibrillation (STCD) strategy using defibrillation shocks preceded by Sync P can improve defibrillation efficacy.

Method and results: We explored the STCD effects in 13 isolated rabbit hearts. During VF, a low-voltage gradient (LVG) area was synchronized by Sync P for 0.92 second. For Sync P, optical action potentials (OAPs) adjacent to four pacing electrodes (10 mm apart) were monitored. When one of the electrodes was in the excitable gap, a 5-mA current was administered from all electrodes. A shock was delivered 23 ms after the excitable gap when the LVG area was unexcitable. The effects of STCD was compared to random shocks (C) by evaluating the defibrillation threshold 50% (DFT(50); n = 35 for each) and preshock coupling intervals (n = 208 for STCD, n = 172 for C). Results were as follows. (1) Sync P caused wavefront synchronization as indicated by a decreased number of phase singularity points (P < 0.0001) and reduced spatial dispersion of VF cycle length (P < 0.01). (2) STCD decreased DFT(50) by 10.3% (P < 0.05). (3) The successful shocks showed shorter preshock coupling intervals (CI; P < 0.05) and a higher proportion of unexcitable shock at the LVG area (P < 0.001) than failed shocks. STCD showed shorter CIs (P < 0.05) and a higher unexcitable shock rate at LVG area (P < 0.05) than C.

Conclusion: STCD improves defibrillation efficacy by synchronizing VF activations and increasing probability of shock delivery to the unexcitable LVG area.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Animals
  • Electric Countershock / methods*
  • Heart Conduction System / physiopathology*
  • In Vitro Techniques
  • Rabbits
  • Treatment Outcome
  • Ventricular Fibrillation / diagnosis
  • Ventricular Fibrillation / physiopathology*
  • Ventricular Fibrillation / therapy*