Rapid left-to-right shunt quantification in children by phase-contrast magnetic resonance imaging combined with sensitivity encoding (SENSE)

Circulation. 2003 Sep 16;108(11):1355-61. doi: 10.1161/01.CIR.0000087603.97036.C2. Epub 2003 Aug 25.

Abstract

Background: Parallel imaging by sensitivity encoding (SENSE) may considerably reduce scan time in MRI. For rapid flow quantification in children with congenital heart disease, we evaluated phase-contrast MRI (PC-MRI) techniques combined with SENSE.

Methods and results: In 22 pediatric patients (mean age, 7.2+/-6.2 years) with cardiac left-to-right shunt, blood flow rate in the pulmonary artery (Qp) and ascending aorta (Qs) and flow ratio Qp/Qs were determined by PC-MRI with SENSE reduction-factor 2 and 3 (SF-2 and SF-3). Additionally, we used PC-MRI with higher spatial in-plane resolution (1.6x2.1 versus 2.3x3.1 mm) with and without SF-3. Results were compared with a recently validated standard PC-MRI protocol and tested in vitro using a pulsatile flow phantom. Reduction of signal averages from 2 to 1 and application of SENSE accelerated flow measurements by a factor of 3.5 (5.2) using PC-MRI with SF-2 (SF-3) compared with standard PC-MRI. For blood flow rate through the pulmonary artery and aorta, as well as for the Qp/Qs ratio we found negligible differences of +/-3%, lower limits of agreement (mean+/-2 SD) of -7% to -18%, and upper limits of agreement (mean+/-2 SD) of +3 to +24%, demonstrating good agreement with standard PC-MRI. Mean Qp/Qs ratio by standard PC-MRI was 1.69+/-0.45 (range, 1.27 to 2.79). Interobserver variability was low, and high accuracy was confirmed in vitro for all protocols.

Conclusions: PC-MRI for flow quantitation may be combined with SENSE to achieve a substantive reduction of scanning time. In children with left-to-right shunt, Qp/Qs quantification is possible by PC-MRI+SF-3 in <60 seconds. Use of higher in-plane resolution did not improve measurement results.

Publication types

  • Clinical Trial
  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Child
  • Child, Preschool
  • Coronary Circulation
  • Female
  • Heart Defects, Congenital / diagnosis*
  • Humans
  • Infant
  • Magnetic Resonance Imaging, Cine / instrumentation*
  • Magnetic Resonance Imaging, Cine / methods
  • Magnetic Resonance Imaging, Cine / standards
  • Male
  • Observer Variation
  • Reference Standards
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Time Factors