Tissue attachment strength of prosthetic meshes used in ventral and incisional hernia repair. A study in the New Zealand White rabbit adhesion model

Surg Endosc. 2002 Nov;16(11):1542-6. doi: 10.1007/s00464-001-8271-y. Epub 2002 Jul 8.

Abstract

Background: Many prosthetic materials are used in incisional hernia repair, including polypropylene (PP) and expanded polytetrafluoroethylene (ePTFE). However, PP forms severe adhesions and ePTFE has raised concerns about the adequacy of tissue attachment.

Methods: The early tissue attachment strength of PP and two new forms of ePTFE (DLM and DLMC) was compared in a rabbit model (n = 12) in which disks of the three meshes (n = 8 of each material) were implanted against the abdominal wall for 3 days.

Results: Tensiometer testing found that DLMC mesh had significantly greater attachment strength than PP (p = 0.02). Histologic studies indicated that this was due to cellular ingrowth. Tissue adhesions were observed with all eight PP disks, one DLMC disk, and none of the DLM disks.

Conclusion: Modified forms of ePTFE mesh may provide abdominal wall repairs that are as strong or stronger than those obtained with PP, with early tissue attachment and without adhesions.

Publication types

  • Comparative Study

MeSH terms

  • Animals
  • Disease Models, Animal*
  • Hernia, Ventral / surgery*
  • Herniorrhaphy*
  • Peritoneum / surgery
  • Polypropylenes / adverse effects
  • Polypropylenes / therapeutic use
  • Polytetrafluoroethylene / adverse effects
  • Polytetrafluoroethylene / therapeutic use
  • Postoperative Complications / surgery*
  • Prostheses and Implants*
  • Rabbits
  • Surgical Mesh*
  • Tissue Adhesions / etiology*

Substances

  • Polypropylenes
  • Polytetrafluoroethylene