The high prevalence of atrial fibrillation (AF) and its clinical complications, the poor efficacy of medical therapy for preventing recurrences, and dissatisfaction with alternative modes of therapy stimulated interest in implantable atrial and combined atrioventricular defibrillators. In a multicenter study, the safety and efficacy of a stand alone implantable atrial defibrillator, the Metrix system, were evaluated. The device was implanted in 51 patients with highly symptomatic episodes of AF refractory to pharmacological treatment. During a follow-up of 9 months, 96% of 227 spontaneous AF episodes were successfully converted to sinus rhythm in 41 patients. In 62 episodes (27%), several shocks and/or additional drug treatment were required to maintain stable sinus rhythm because of early recurrences of AF. A total of 3719 shocks were delivered and no induction of ventricular proarrhythmia or inaccurately synchronized shocks occurred. The AF detection algorithm exhibited a 100% specificity for the recognition of sinus rhythm and a 92.3% sensitivity for the detection of AF. The combined atrioventricular defibrillator, Jewel AF 7250, was evaluated in a multicenter, randomized, cross-over trial. The primary study objectives included: overall safety as determined by complications-free survival at 6 months, efficacy of tiered atrial pacing and defibrillation therapies for termination of spontaneous atrial tachycardias (AT) and AF, and relative sensitivity of a new dual-chamber detection algorithm. The device was implanted in 211 patients with either a history of ventricular tachyarrhythmias (VT/VF) alone or with a history of both AT/AF and VT/VF. During a mean follow-up of 4.5 months, it has been shown that the Jewel AF is safe and effective in treating atrial and ventricular tachyarrhythmias. Pace termination of 85% of AT episodes were achieved with painless delivery of antitachycardia pacing; additional 35% of AT episodes were terminated by high frequency burst pacing.
Conclusions: The stand alone implantable atrial defibrillator may be safe and clinically useful in selected patients for the treatment of highly symptomatic, drug resistant recurrences of AF. The combined atrioventricular defibrillator may be particularly indicated in patients presenting with both a history of atrial and ventricular tachyarrhythmias.