Objective: The purpose of this study was to determine if the atrial response upon cessation of ventricular pacing associated with 1:1 ventriculoatrial conduction during paroxysmal supraventricular tachycardia is a useful diagnostic maneuver in the electrophysiology laboratory.
Background: Despite various maneuvers, it can be difficult to differentiate atrial tachycardia from other forms of paroxysmal supraventricular tachycardia.
Methods: The response upon cessation of ventricular pacing associated with 1:1 ventriculoatrial conduction was studied during four types of tachycardia: 1) atrioventricular nodal reentry (n = 102), 2) orthodromic reciprocating tachycardia (n = 43), 3) atrial tachycardia (n = 19) and 4) atrial tachycardia simulated by demand atrial pacing in patients with inducible atrioventricular nodal reentry or orthodromic reciprocating tachycardia (n = 32). The electrogram sequence upon cessation of ventricular pacing was, categorized as "atrial-ventricular" (A-V) or "atrial-atrial-ventricular" (A-A-V).
Results: The A-V response was observed in all cases of atrioventricular nodal reentrant and orthodromic reciprocating tachycardia. In contrast, the A-A-V response was observed in all cases of atrial tachycardia and simulated atrial tachycardia, even in the presence of dual atrioventricular nodal pathways or a concealed accessory atrioventricular pathway.
Conclusions: In conclusion, an A-A-V response upon cessation of ventricular pacing associated with 1:1 ventriculoatrial conduction is highly sensitive and specific for the identification of atrial tachycardia in the electrophysiology laboratory.